XZ-P6 vacuum blood collection tube special capping centrifuge manual

First, the product model, name:

XZ-P6 vacuum blood collection tube special capping centrifuge

Second, use and scope of use:

The XZ-P6 vacuum blood collection tube special capping centrifuge (hereinafter referred to as the machine) can be widely used in clinical medicine, biochemistry, immunology and other fields, and is a conventional instrument for centrifugal sedimentation in the laboratory.

Third, the main specifications and technical parameters

Maximum speed 4000r/min
Maximum centrifugal acceleration 3350×g
Timing range 0~99min
Speed ​​display accuracy ±30r/rpm
Noise 65dB
Appearance size 600 × 490 × 370mm
Weight 45kg

Rotor speed and centrifugal force parameters:

Rotor number

Maximum speed

Maximum capacity

Maximum centrifugal force

No.1 horizontal rotor

4000r/min

72x7ml/5ml

3350xg

No.2 horizontal rotor

4000r/min

80x7ml/5ml

3350xg

Fourth, the control panel function introduction


Xz-p6 capping centrifuge control panel
1, select the settings you need
2, increase the button
3, function parameter drop button
4, confirm the function parameter confirmation button
5, start the centrifuge speed button
6, centrifuge speed stop / door lock button
7, jog keys
8, speed and centrifugal force display switch button
9, display failure display:
E1=Unbalance E2=Overspeed E3=Door switch (start)
E5=Time setting is zero E6=Temperature
E7=Motor does not turn E8=Communication E9=System

Fifth, the operation process

Operation procedure (1) Turn on the power, press the door lock button to open the door cover automatically; the centrifuge tube in the centrifugal sample loading area is placed symmetrically in the rotor, and each tube sample can be visually observed.
(2) Press the Setup button until the desired item window flashes, press the +, - buttons to set its contents, and then press the Enter key.
(3) Parameter setting Press the setting button rotor number window to flash, press +, - to select the rotor number;
Press the set button speed/centrifugal force window to flash and press +, - to select the speed;
Press the Setup button to raise/decelerate the window and press the +, - buttons to select the up/down speed;
Press the setup button time window to flash and press the +, - buttons to select the time;
The selected parameters are correct, press the confirmation button, which is the parameter you are currently setting.
(4) When the parameters are selected, press the approval button again. During the operation, the centrifugal force can be directly displayed. Press the speed/centrifugal force switch button. (Note: the speed can still be modified during the operation.) Press the start button and start. jobs. Time to zero or press the stop button, after the speed is zero, press the door lock button to automatically open the door.

Sixth, working principle and characteristics

The machine is directly driven by a motor, and the special damper is used for vibration reduction. The rotor is precision casted with high-strength materials, and the rotor is easy to handle. Microcomputer control, AC variable frequency motor, high precision, low noise. Large screen LCD display, speed, centrifugal force, running time, fault display. The unique design of the double layer of the adapter prevents the cap from falling out during the operation of the centrifuge, and automatically removes the cap to improve the efficiency of the experiment. The rise/deceleration time is short (30S) with 1 to 8 liters/downshift. With electronic door lock, overspeed, automatic rotor identification, unbalanced detection and protection to ensure personal safety. Touch panel, programmable operation, automatic conversion of rotational speed centrifugal force, easy to operate. It is suitable for centrifugation in hospital biochemistry, radiology laboratory, and central blood station biological products.
VII. Installation procedure (1) The table top is solid and flat to prevent the instrument from vibrating or moving the instrument during operation.
(2) The power supply has an independent ground line. It is strictly forbidden to share the neutral line and the ground line to prevent electric shock and injury.
(3) There should be no corrosive gas and electromagnetic field interference around the instrument to prevent premature failure of the external protective layer of the instrument and interference with electrical components.
(4) The instrument is placed horizontally to ensure that the four rubber feet are evenly stressed.
When the time reaches zero or the stop button is pressed, the automatic door cover opens automatically.


Source: Centrifuge

Royal Jelly is also called Bee Milk. The fresh royal jelly is slightly ropy milk paste  substance; it is the excretive mixture of nutrition gland and maxilla gland of the head of little worker bee. The worker bees use this to feed the 1-3 days` worker bee larva and drone larva, 1-5.5 days` queen bee larva and queen bee in the oviposition period. The royal jelly is the biologic product containing very complicated active elements which contains almost all the nutrition elements required by the growth of the human body.